Navigation:
 
 


 
 
MyHuber
Create your own navigation!

400°C 300°C 200°C 100°C -10°C -20°C -30°C -40°C -50°C -60°C -70°C -80°C -90°C -100°C -110°C -120°C

ATEX

Two alternative solutions are available: Using the ATEX conform Unistat remote control II 2G EX ib IIC T4 where the Unistat is located outside the Ex zone OR the Unistat is installed in an Ex-p pressure enclosure within the Ex zone.



Description: Ex-p enclosure for zones 1 and 2 with pressure encapsulation to EN 60079-2
Type: Ex p II T4
Features: Stainless steel construction
 
• Door with Ex double door seal and turnbuckle latch
 
• 1 breakout with safety film for installation of Unistat Pilot (standard operability and functionality are not impaired)
 
• 2 x bulkhead fittings for cooling water connection
 
• Conductive rollers
 
• Ex-p enclosure pressurized encapsulation to EN 60079-2
 
• Ex ia temperature measurement (Cat.No. 9399)
 
• 2 x metal braided hoses for cooling water connection
 
• redundant temperature monitoring
 
For unichillers we can offer a custom quotation.  Please provide us with zone, explosion sub-group and temperature class when requesting information.